computer type Silicon-carbon analyzer (JS-TG6 type) |
|
JS-TG6 type Stokehole Silicon-carbon analyzer has said Stokehole liquid iron quality control instrument that I plant a new generation of products, also known as casting Stokehole Laboratory instruments, Stokehole analyzer, high accuracy, easy to use and reliable. Cast iron Stokehole rapid thermal analysis Cast iron technology is based on the formation of tissue coagulation temperature curve of the measured object by the computer on the solidification temperature curve into line analysis, to calculate the molten iron of carbon equivalent (CE), the carbon content (C%), silicon content (si% ), ball rate (SG%), tensile strength (RM) and other indicators, automatically calculate recarburiser, ferrosilicon, put the amount of scrap iron component is adjusted, Stokehole liquid iron castings quality management instrument is to achieve quality optimization control effective way.
Instrument
Device
Regulation
Grid
|
Product Name
|
Stokehole liquid iron quality management instrument
|
type No.
|
JS-TG6
|
Host computer
|
Military grade Motherboard
|
Working humidity
|
0 ~ 50 ℃
|
Monitor
|
19 inches TFT color display
|
Operating temperature
|
<76%
|
Power supply
|
AC 220V 50Hz
|
Power
|
200W
|
Shape
|
Desktop
|
Detection time
|
90 seconds
|
The main function
|
Measurement function
|
The carbon content of the molten iron grade silicon content of carbon equivalent degree of supercooling rate of the ball tensile strength
|
Setting Function
|
You can set the calibration curve 10, 10 kinds of target components, calibration curve and the target material linkage
|
Fill in the function
|
Based on the measurement results against the target component, iron weight, yield calculation additional amount
|
Record
|
By measuring the time sequence, it can record 5,000 measurements can reproduce process
|
Browsing
|
Visit the main trends in the distribution of the measured parameters, solidification temperature differential curve diagram
|
Output function
|
Optional measuring and recording functionality to U disk automatically output measurement data
|
Measurement accuracy
|
Project name
|
Measuring range
|
Measurement accuracy
|
Resolution
|
0.1 ℃ (0-1370 ℃)
|
0.1% FS
|
CE% value
|
2.30% -4.83%
|
± 0.10%
|
C% value
|
2.20% -4.30%
|
± 0.05%
|
Si% value
|
0.30% -3.80%
|
± 0.10%
|
SG% value
|
60% -95%
|
± 10.0%
|
RM value
|
HT100-HT350
|
Min
|
|
|
|